Cookies sur les sites Web Omega
Nous utilisons des cookies sur ce site Web. Ces cookies sont indispensables au bon fonctionnement de notre site.
Si vous continuez sans modifier vos paramètres, nous supposerons que vous êtes heureux de recevoir tous les cookies de notre site Web.
Pour de plus amples informations à ce sujet, veuillez cliquer ici
Fermer
Panier  |  Contact  | 
NUMÉRO VERT0805 541 038
INTERNATIONAL+33 01 57 32 48 17

Votre unique source pour la mesure et le Contrôle de Procédés

RTD Input Non-Isolated DIN Rail Loop-Powered Signal Conditioner

DRSL-RTD-LP

RTD Input Non-Isolated DIN Rail Loop-Powered Signal Conditioner | DRSL-RTD-LP
Nouveau

70,00 DRSL-RTD-LP

Garantie 1 an CE
  • 2, 3 or 4-Wire Pt100 RTD Input
  • Powered By the Host (Output) Current Signal Loop
  • High Accuracy Better Than 0.2°C or 0.1% of Selected Range
  • Slimline 6 mm (0.24") Housing
  • Excellent EMC Performance and 50/60 Hz Noise Suppression
  • Fast Response Time < 30 ms/300 ms (Selectable)
  • Pre-Calibrated Temperature Ranges Selectable via DIP-Switches
Instruments de mesure et contrôle de processus - Voir Les Produits Associés

Déscription

The DRSL-RTD-LP RTD input non-isolated DIN rail loop-powered signal conditioner measures a standard 2-, 3- or 4-wire Pt100 temperature sensor and provides a passive analog current output signal. The DRSL-RTD-LP provides a competitive choice in terms of both price and technology for interfacing RTD signals to SCADA systems or PLC equipment. This unit is powered by the host (output) current signal loop. Low power consumption facilitates DIN rail mounting without the need for any air gap. Easy configuration of more than 1000 factory calibrated measurement ranges is done via DIP-switches. The unit operates over a wide temperature range from -25 to 70°C (-13 to 158°F).

SPECIFICATIONS
INPUT
Input Type: 2, 3, or 4-wire Pt100 RTD
Temperature Range: -200 to 850°C (-328 to 1562°F)
Sensor Current, RTD: <150 µA
Sensor Cable Specifications: 50Ω per wire or 50 nF
Effect of Sensor Cable Resistance (3 or 4-Wire RTD): <0.002Ω/Ω
Broken Sensor Detection: >800Ω
Shorted Sensor Detection: <18Ω

OUTPUT
Current Output
Programmable Signal Ranges: 0 to 20 mA and 4 to 20 mA
Range Limits (NAMUR NE43 Out of Range): Below 3.8 mA or above 20.5 mA
Sensor Error Detection (Dip Switch Selectable for Enable or None): Below 3.5 mA or above 23 mA
Incorrect DIP-Switch Setting Identification: Below 3.5 mA or above 23 mA
Output Error Level: DIP switch selectable for upscale or downscale
Load Resistance (Ω): ≤(Vsupply -3.3)/ 0.023
Load Stability: ≤0.01% of span/100Ω

GENERAL
Supply Voltage: 3.3 to 35 Vdc
Voltage Drop: 3.3 Vdc
Power Consumption: 1 W max
Internal Consumption: 0.65 W max
Signal/Noise Ratio: >60 dB
Response Time (0 to 90%, 100 to 10%): <30 ms/300 ms (selectable, provides either fast response or signal dampening as needed).
Accuracy: Better than 0.2°C or ±0.1% of selected range
Temperature Coefficient: ≤±0.02°C/°C
EMC Immunity Influence: <±0.5% of span
Extended EMC Immunity
    NAMUR NE 21, A Criterion, Burst: <±1% of span (span = selected input range)

ENVIRONMENTAL
Operating Temperature: -25 to 70°C (-13 to 158°F)
Storage Temperature: -40 to 85°C (-40 to 185°F)
Calibration Temperature: 20 to 28°C (68 to 82°F)
Relative Humidity: 0 to 95% RH non-condensing
Protection Degree: IP20
Installation Area: Pollution degree 2 and measurement/overvoltage category II

MECHANICAL
Dimensions: 113 H x 6.1 W x 115 mm D (4.4 x 0.24 x 4.5")
Weight: 70 g (0.15 lb) approx
DIN Rail Type: DIN EN 60715 - 35 mm
Wire Size: 0.13 x 2.5 mm²/AWG 26 to 12 stranded wire
Screw Terminal Torque: 0.5 Nm
Placez votre commande(Veuillez spécifier la référence du produit)
Code d'article/ Desc.
Stock
Prix
Déscription
Qté
DRSL-RTD-LP
Star
Stock: 6 semaines  
70,00   
RTD input non-isolated DIN rail loop-powered signal conditioner
 
Accessories
RAIL-35-1
Star
Stock: 1 semaine  
8,80   
35 mm (1.4") DIN rail, 1 m (3.3') length
 
DRSL-MOD-STOP
Star
Stock: 9 semaines  
4,00   
Module stop (screwed onto DIN rail to support and hold mounted devices)
 
Tous les Montants indiqués Dans EUR, GBP, USD.
Exemple de commande: (1) DRSL-MOD-STOP Module stop (screwed onto DIN rail to support and hold mounted devices), €4,00  plus (1) DRSL-RTD-LP RTD input non-isolated DIN rail loop-powered signal conditioner, €70,00  plus (1) RAIL-35-1 35 mm (1.4, €8,80, €4,00 + 70,00 + 8,80 = €82,80

Manuels de Produit:

Télécharger Installation Guide for Signal Conditioners
Télécharger DIN Rail Isolators and Signal Conditioners